W203 2.0 Electrolytic Sensor Method Water Vapour Permeability Analyser

Product Introduction

Based on the test principle of electrolytic sensor method, the W203 2.0 Water Vapour Permeability Analyser is designed and manufactured according to GB/T 21529, ISO15106-3 standard, based on the test principle of electrolytic method. W203 2.0 is suitable for the water vapour transmission performance test of films, sheets, papers, packages and various materials in the fields of food, medicine, medical equipment, daily chemicals, photovoltaic electronics, etc. It is an ideal configuration instrument for testing the barrier properties of packaging materials in the packaging production industry whether offline or online.

Applied to:

  • Plastic film, composite film, aluminium foil, aluminised film, etc.
  • Sheet, panel, rubber, ceramics, etc.
  • Packaging containers, such as: glass, bottle, cans, boxes, etc.
  • Expanding application: solar panel, LCD film, medical patch, etc. Widely used in quality inspection organisations, drug control institutions, research institutes, packaging, thin film, food companies, pharmaceutical enterprises, personal care industry, electronics industry and so on.

 

Standards

ISO15106-3, ASTM E398, GB/T 21529, DIN 53122-2, YBB00092003

 

Features

 

Simple Operation

Professional software with a simple interface, easy to use and convenient to set test process. Fully-auto operation, one-button test, judge and stop
automatically. Mainframe configured with colour touch screen, can observe temperature, humidity and transmission without external computer. Professional test report can be automatically generated and exported in PDF format.

Advanced Technology

Humidity control: Dual gas flow method and constant humidity. Supports electronic signature, online report submission and auditing function.

High Efficiency

With three different test modes of high, medium and lower barriers, it can test films with different barrier property. Measurement precise up to 0.001 g/m2·24h, can test high barrier materials, such as aluminium foil.

Authority management & data trace

The software is designed according to the requirements of the new GMP Appendix computerised system. You need a user name and password to log in to the workstation, and ensure the safety and effectiveness of account and experiment data. Users can be classified into different levels of system administrator, instrument administrator, auditor and operator. The system administrator can adjust the permissions of various levels, for example, increase and decrease system control items of any level. Audit and trace functions (test tracking, log tracking) are available so every data change is recorded, ensuring the security and integrity of test data.

Reliable and easy-maintenance instrument

Imported infrared sensor, with high precision and good performance which can work for a long time. Sensor over-range automatic protection prevents damaging important sensors. Functional modular design and easy to maintain.

 

Specifications

Item Technical Parameters
Test Range 0.001~40 g/(m² .24h) (film and sheet)
Test Precision 0.0001 g/(m² .24h) (film and sheet)
Temperature Range 15~45°C
Temperature Accuracy ±0.1°C
Humidity Range (5~90) %RH
Humidity Accuracy ±1%RH
Test Area 50.24 cm²
Sample Size Φ100 mm
Sample Thickness ≤3 mm
Number of Test Samples 3 Pieces
Carrier Gas 99.999% Nitrogen (user provide)
Carrier Gas Pressure ≥0.1 MPa
Carrier Gas Flow 0~100 mL/min
Pneumatic Pressure ≥0.1 MPa
Instrument Size 670×550×400mm
Weight 57kg
Power 750 W
Power Supply AC 220 V,50 Hz (110V optional)

 

Downloads

W203 2.0 Brochure
PDF – 406.1 KB


Get in touch to discuss your requirements